Controllers
12Páginas

{{requestButtons}}

Extractos del catálogo

Controllers - 1

sc1000 and sc100 Digital ControllersՕ NEW! si792 2-Wire TransmitterModel 53 and 33 Controllers ՕPRO Transmitter >

Abrir la página 1 del catálogo
Controllers - 2

Multi-Parameter, Modular, Multi-Parameter > Controllers Digital / Analog Controller Digital Controller w/ Touch Screen Interface Parameters Multi-parameter: See pages 397-467 for details.Multi-parameter: See pages 397-467 for details. Display Graphic Backlit LCDBacklit LCD User Interface Touchscreen / Ethernet / WirelessKeypad Number of Plug-and-Play 2 to 8 Sensors or up to 2 Analyzers; 2 Sensors* Digital Sensor Inputs maximum of 8 digital sensor/analyzer combinations per sc1000 probe Module* Number of 0/4-20 mA Analog Inputs Expandable to 12 None 0/4-20 mA Isolated Outputs Expandable to...

Abrir la página 2 del catálogo
Controllers - 3

Full-featured, 2-wireRobustPanel Mounted2, 3, or 4-Wire Analog TransmitterAnalog ControllerAnalog ControllerAnalog Transmitter pH, conductivitypH, conductivity, DO, FlowpH, conductivity, DO, FlowpH, conductivity, DO, Flow LCDBacklit LCDBacklit LCDLCDKeypadKeypadKeypadKeypad NoneNoneNoneNone NoneNoneNoneNone 1221None42None NoneNoneNoneNoneHart > , PROFIBUSPA,Hart > ή Hart > None Foundation FIELDBUS NoneNoneNoneNonePanel, Surface, or PipePanel, Surface, or PipePanel Panel, Surface, Pipe, or Integral Sensor 14.4 x 14.4 x 10.5 cm14.4 x 14.4 x 15.0 cm10.5 x 9.0 x 12.5 cm9.5 x 9.5 x 6.0 cm (5.67...

Abrir la página 3 del catálogo
Controllers - 4

The Hach sc1000 Multi-parameter Universal Controller is a fullymodular system consisting of a Display Module and one or more Probe Modules. > The sc1000 Display Module is a full-featured controller with a large color touch-screen display. The intuitive, easy to use interface can be used for any number of parameters. One Display Module controls either a single Probe Module or a number of Probe Modules connected by a digital network. The Display Module is fully portable and can be disconnected and moved anywhere within the network. > Controllers Each sc1000 Probe Module provides power to the...

Abrir la página 4 del catálogo
Controllers - 5

Controllers The sc1000 Display Module allows the user not only to list ameasurement value but also to show a trendline of the measurement by day or week. Additionally, the Display Module has an interactive chart page for analyzing any number of trendlines. There are a number of ways to access the information from an sc1000 Network. The standard 0/4-20 mA and BUS networks are available through the sc1000 Probe Module. The sc1000 Display Module has a dedicated Ethernet service port for remote access or access via PC or laptop. A final option is Wireless GSM Telecommunications offered as a...

Abrir la página 5 del catálogo
Controllers - 6

One Controller for One or Two Sensors > The Hach sc100 Controller receives data from up to two sensors.Use any of Hachْs line of digital sensors for a wide variety of parameters. One Controller for One or Two Parameters > Not only can the sc100 controller be used for up to two sensors, but the sensors need not be the same. Mix and match any combination of parameters. Controllers Plug-and-Play Operation > Theres no complicated wiring or set up procedures with the sc100 controller. Just plug the sensor in and itҒs ready for use without special ordering or software configuration. Flexible...

Abrir la página 6 del catálogo
Controllers - 7

Controllers Alarms Low alarm point, low alarm pointdeadband, high alarm point, high alarm point deadband, off delay, and on delay Communications (Standard) Datalog access and software upgradeaccess via standard RS-232 service port and wireless infrared port Communications (Optional) RS-232 (MODBUS): Configure andretrieve measured data for oneanalyzer using IBM-compatible PC RS-485 (MODBUS) & PROFIBUSDPV1: Advanced communications/networking with PLC or SCADAsystem directly from analyzer. Memory Backup All user settings are retainedindefinitely in memory (non-volatile) (EEPROM) Mounting...

Abrir la página 7 del catálogo
Controllers - 9

Controllers Separate passwords can be set for the following: Sensor Calibration Օ Transmitter Configuration Administrator > The transmitterՒs electronics are attached to the hinged door and are protected against aggressive environments. The easy-to- access and clearly described terminals ensure comfortable and error-free wiring.The empty rear enclosure can be pre-mounted, which makes the wiring and installation of the instrument (Panel, Wall or Pipe Mounting) very easy. The door can be attached afterwards with the simple, robust hinge. > Easy-access, clearly-described terminals ensure...

Abrir la página 9 del catálogo
Controllers - 10

Surface, panel, and pipe (horizontal and vertical) > Enclosure Ambient Conditions NEMA 4X/IP66; metal enclosure with corrosion-resistant finish Dimensions Ultrasonic Flow , U53 Operational: -4 to 140F (-20 to 60ںC); 0 to 95% relative humidity, non-condensingStorage: -22 to 158F (-30 to 70ںC); 0 to 95% relative humidity, non-condensing 1/2 DIN; 144 x 144 x 150 mm (5.7 x 5.7 x 5.9 in.) Flow rate: 0-9999, 0-999.9 or 0-99.99 with selectableflow rate units and multiplier Volume: 0-9,999,999 with selectable volume unitsDepth: 0-1200.0 inches, 0-100.0 feet, 0-30,000mm, or 0-30.00 meters Weight 1.6...

Abrir la página 10 del catálogo
Controllers - 11

Enclosure Accuracy (Controller) Polycarbonate enclosure with IP65 front panel;general purpose; two brackets for panel mounting P33, C33, F33: 0.1% of spanD33: 0.1% of span E33: 0.5% of span Ambient Conditions Operational: -4 to 140ѺF (-20 to 60C); 0 to 95%relative humidity, non-condensingStorage: -22 to 158ںF (-30 to + 70 C); 0 to 95%relative humidity, non-condensing Stability Dimensions P33, C33: 0.05% of span per 24 hours., non-cumulativeE33: 0.2% of span per 24 hours., non-cumulative 1/4 DIN; 94 x 94 x 142 mm (3.7 x 3.7 x 5.6 in.) Weight Power Requirements 0.8 kg (1.7 lbs.) Repeatability...

Abrir la página 11 del catálogo
Controllers - 12

Stability PRO-P3, PRO-C3: 0.05% of span per 24 hours., non-cumulativePRO-E3: 0.2% of span per 24 hours., non-cumulative > Weight 10 oz. (280 g) approximately Repeatability Ambient Conditions PRO-P3, PRO-C3, PRO-E3: 0.1% of span or betterPRO-D3, PRO-F3: ձ0.05% of span Operational: -4 to 140F (-20 to 60ںC); 0 to 95% relative humidity, non-condensingStorage: -22 to 158F (-30 to 70ںC); 0 to 95% relative humidity, non-condensing Certifications CE, Class 1 Div II Class A, B, C, D Hazardous locations Measurement Temperature Drift Power Requirements pH/ (ORP), PRO-P3 PRO-P3, PRO-C3: Zero & Span: <...

Abrir la página 12 del catálogo

Todos los catálogos y folletos técnicos Hach

  1. Flow

    22 Páginas

  2. Process (4)

    14 Páginas

  3. Process (3)

    18 Páginas

  4. Process (2)

    24 Páginas

  5. Process (1)

    16 Páginas

  6. Lab Supplies (4)

    16 Páginas

  7. Lab Supplies (3)

    16 Páginas

  8. Lab Supplies (2)

    18 Páginas

  9. Lab Supplies (1)

    16 Páginas

  10. Parameters (3)

    26 Páginas

  11. Parameters (2)

    24 Páginas

  12. Parameters (1)

    28 Páginas

  13. Sample Cells

    8 Páginas

  14. Lab Turbidimeters

    6 Páginas

  15. Meters & Probes

    26 Páginas